鞋类防滑性能测试方法的重复性与再现性研究

Study on the Repeatability and Reproducibility of Test Methods for Footwear Slip Resistance

  • 摘要:
    目的 文章旨在完善鞋类防滑性能测试方法的精密度数据,支撑国家标准修订与实验室质量控制。
    方法 依据GB/T 6379.2开展鞋类防滑性能测试方法的精密度研究,组织14家实验室对成品鞋和片材样品开展了比对。采用Mandel hk统计量进行一致性检验,Cochran检验和Grubbs检验识别离群值。
    结果 比对的实验室整体测试能力良好,防滑测试方法的重复性良好,再现性在干湿条件下差异显著。
    结论 研究获得三类样品在干湿条件下的系统性精密度数据,可为实验室日常质量控制和测量不确定度评定提供参考,也为GB/T 3903.6的标准修订提供关键技术依据。

     

    Abstract:
    Objective The slip resistance of footwear is a critical performance indicator for ensuring the walking safety of wearers and preventing slip-related accidents. Most domestic footwear product standards adopt GB/T 3903.6 whole shoe slip resistance performance test method standard as a reference for specifying slip resistance requirements. However, the absence of a precision clause in this method standard compromises the inter-laboratory comparability of test results and weakens the basis for conformity assessment in product standards. This study aims to evaluate the repeatability and reproducibility of GB/T 3903.6, thereby providing reliable data to support the development of a precision clause for the standard and to facilitate laboratory quality control.
    Methods An interlaboratory collaborative study was conducted involving 14 laboratories from across China in accordance with GB/T 6379.2. The test specimens included finished footwear, sheet materials, and S96 standard rubber. A total of 14 test levels were established, comprising 8 levels for finished footwear, 4 levels for sheet materials, and 2 levels for S96 standard rubber, with 5 replicate measurements at each level. Mandel h and k statistics were adopted for consistency evaluation, while Cochran and Grubbs tests were used for outlier identification. Based on the one-way analysis of variance model, the repeatability standard deviation (Sr), reproducibility standard deviation (SR), as well as their corresponding coefficients of variation (CVr and CVR) were calculated.
    Results (1) Consistency analysis revealed that one laboratory (Lab 12) exhibited systematic positive bias across all 12 test levels, with Mandel h values exceeding the 5% critical value (1.85) in 8 levels. This laboratory was excluded after confirmation by the Grubbs test, leaving 13 valid laboratories for subsequent analysis. (2) The Sr values of all test levels ranged from 0.007 to 0.010, with CVr ranging from 0.8% to 3.2%, indicating consistently low and stable within-laboratory variability and demonstrating good repeatability. (3) The reproducibility coefficient of variation CVR differed markedly between dry and wet conditions. Under dry conditions, the CVR values for finished footwear, sheet materials, and S96 standard rubber were 9.7%-15.4%, 6.8%-7.6%, and 3.8%, respectively; under wet conditions, these values increased to 23.9%-34.6%, 11.9%-12.5%, and 10.9%, respectively. Notably, finished footwear exhibited lower absolute SR values under wet conditions (0.064-0.092) than under dry conditions (0.087-0.132); the elevated CVR was mainly attributed to the substantially lower mean friction coefficients under wet conditions, which amplified the relative variability. (4) Under identical test surface conditions, CVR increased in the order of standard rubber < sheet materials < finished footwear, demonstrating that differences in specimen homogeneity and shoe last assembly were important sources of inter-laboratory variability. (5) Compared with the international precision study by Jung and Fischer (1993), the wet-condition CVR values in this study (23.9%-34.6%) were consistent with the historical data (25%-36%), indicating that the reproducibility performance of this test method has remained relatively stable over the past three decades.
    Conclusions The GB/T 3903.6 method demonstrates good repeatability, but its reproducibility differs significantly between dry and wet conditions, with markedly elevated CVR observed for wet-condition tests and finished-footwear specimens. The systematic precision data reported in this study, covering three specimen types across 14 test levels, can be applied to routine laboratory quality control and measurement uncertainty evaluation. Furthermore, the identified influences of surface condition and specimen assembly complexity on reproducibility provide a useful reference for unifying relevant test parameters and for developing a precision clause in the future revision of GB/T 3903.6.

     

/